Specifications

4-6
Operating Features and Specifications
Data and Data (inverted) Outputs
Data and Data (inverted) Outputs
The following pattern generator data output features are accessed using the
key and its menu of softkeys.
Features
Polarity - normal or inverted data.
Data high level adjust.
Data amplitude adjust.
External termination voltage 0/2V or ac coupled.
External attenuator set 0 to 40 dB for 0V termination.
Delay v clock adjust.
Output ON/OFF selection.
Vertical data-eye cross-over adjust.
Independent control of high level, amplitude and ON/OFF for DATA and DATA.
Data output features not controlled by data output softkeys
Output gating - controlled by AUXILIARY INPUT port (see Auxiliary Input descrip-
tion on page 4-10).
Error add - (see Error Add paragraph on page 4-8).
Specifications
Data outputs - main
Interface: complementary dc coupled, reverse terminated.
Impedance: 50nominal.
Format: NRZ, normal or inverted.
Amplitude: 0.5 to 2 V pp in 10 mV steps.
Transition times: (10% to 90%) < 30 ps (typical at 2 V pp).
Jitter: typically less than 20 ps pp; <15 ps pp typical at 10 Gb/s
Offset (range): +1.5 V to 3.0 V in 10 mV steps.
Clock/data delay: ±1 ns in 1 ps steps (100 MHz to 500 MHz),
1 clock period (500 MHz to 12.5 GHz).
Resolution: ±1ps.
Connector: APC-3.5 male connector.
Terminations: 50Ω−2V; 50ac coupled
data output