Specifications
2-19
Installation
Selftest at Power-on
Selftest at Power-on
At power-on the error performance analyzer system performs a selftest (this takes
approximately 15 seconds to complete). During this time the display, instrument, clock
source and mainframe (option UKB instruments) operate as follows:
Display: The display is blank for the first few seconds of the selftest. It
then shows a multi-colored raster. The raster sweeps to the right,
to show a blue back-ground. For the remainder of the selftest the
display is as follows:
After selftest the display may continue to display the above, or
will display the module parameters present prior to the last power
down.
Agilent 70843: All front panel indicators are lit for approximately eight seconds
then extinguished for the remainder of the selftest.
After selftest the ACT indicator should light.
Clock Source Module: All front panel indicators are lit for approximately five seconds
then extinguished for the remainder of the selftest.