Specifications

2-16
Installation
System Installation
Procedure
Use the following procedure to install your Agilent 71612 series system.
CAUTION Ensure that no power cables are connected. Also check that the LINE
POWER switches are set to OFF.
CAUTION Ensure that the display line voltage selector switches are set for the line
voltage being used, also check the fuse ratings, see pages 2-8 and 2-10.
1. Install the Agilent 70340A clock source module into the display. Refer to Installing an
Agilent 70340A Clock Source Module into a Display on page 2-20.
2. Arrange the Display and Agilent 70843 for bench operation. The plastic feet on the
Display and Agilent 70843 are self-aligning when systems are stacked. To rack mount
your system, refer to Rack Mount Installation, see page 2-13.
3. Connect the HP-MSIB cables as follows:
CAUTION Your system must be powered down when connecting or disconnecting
HP-MSIB cables.
The diagram shows the systems viewed from the rear.
4. Connect the CLOCK IN port of the Agilent 70843 pattern generator to the CLOCK
OUT of the clock source module.
CAUTION When tightening SMA connectors and terminations, ensure that
the maximum torque setting used is 0.9 N-m.