Specifications
2-13
Installation
Preparation for Use
If you want to change the GPIB address (for example, use an address that is different from
that defined by the column switch settings), it is recommended that you use the Display,
Address Map function keys as follows:
CAUTION It is not recommended that you change the GPIB address using the HP-
MSIB/GPIB switches as these also change the HP-MSIB address. If
the HP-MSIB address protocol is violated your system will fail to
operate.
To Change the GPIB Address
1. Press the Display fixed label key.
2. Press the left-menu softkey.
3. Rotate the large display knob clockwise until the green box rests on the 70843 Err Perf
Anl.
4. Press , select an address using the numeric keypad then press
.
Factory Preset GPIB Addresses
The error performance analyzer GPIB address is factory preset to 18 (column part of
HP-MSIB switch setting).
Bench Operation
Plastic feet are included with Mainframes and stand-alone instruments to provide bench
operation convenience. The plastic feet are self-aligning when systems are to be stacked.
Rack Mount Installation
Front handles must be removed when fitting the system rack mount options.
Agilent 71612 option 1CM - rack mount kit (part number 15810A)
Agilent 71612 option 1CP - rack mount kit with handles (part number 15811A)
Agilent 70843 option 1CM - rack mount kit (part number 15810-60001)
Agilent 70843 option 1CP - rack mount kit with handles (part number 15811-60001)
DISPLAY
Address Map
HP-IB ADDRSET
ENTER