Specifications
5-53
System Command Reference Section
IEEE Common Commands and Queries
Effect of Common Command *RST
The *RST command sets the device-specific functions to a known state that is
independent of the past-use history of the device. The command has the same effect as the
front-panel key.
In addition, receipt of *RST by the error detector will cause all past results to be labelled
invalid.
NOTE The GP-IB bus will be held off for approximately 8 seconds following
receipt of this command to allow the instrument to settle fully.
Scope of the Common Command *TST?
The instrument settings are not affected by this command. This command is not accepted
by the error detector if it is currently gating.
The scope of this command is to test the processor RAM, ROM, DIL switches, gate array
RAM, dual-port RAM, measurement processor and floppy disk drive
Response to Identification Common Query *OPT?
Two comma-separated fields are returned:
Effect of the Common Command *RCL
Syntax “*RCL <NRf>”
This command recalls the setup from a numbered store. The range of store numbers is 0
through 9.
In addition, receipt of *RCL by the Error detector will cause all past results to be labelled
invalid.
NOTE The GP-IB bus will be held off for approximately 8 seconds following
receipt of this command to allow the instrument to settle fully.
Field one: UHF - Error Detector and Pattern Generator.
UHH - Error Detector only.
UHG - Pattern Generator only.
Field two UHJ - Error Location option installed.
0 - Error Location option not installed.
INST PRESET