Specifications
5-45
System Command Reference Section
Error Detector Result and Configuration Logging
Error Detector Result and Configuration Logging
The logging of error detector results and configuration settings to an GP-IB printer at
address 1 is mutually exclusive with any remote control of the instrument using the same
GP-IB interface.
Although the logging of error detector results and configuration settings may
predominantly be of use when the instrument is under local control and configured with
the “Log to” setting configured to either “GPIB prnter,addr 1” or “RS232 printer” the
logging capabilities may also be used in conjunction with or instead of the query
commands defined in Error Detector Measurement Functions on page 5-36.
While under remote control via the rear panel GP-IB interface the SENSe1:LOGG:PORT
command does, however, permit redirection of the logging information between a printer
connected to the RS232 interface and the controlling computer system.
The SENSe1:LOGG commands control when a line of text is generated and made
available to be read by the controller. An SRQ is asserted when a line of text is available.
Refer to FETCH[:SENSe[1]] :LTEXt? on page 5-39 for a definition of the
command which provides remote querying of the logged information.
SENSe[1]:LOGGing ONCE|<boolean>
The LOGGing ONCE command is equivalent to the front-panel key.
The LOGGing <boolean> command enables and disables the logging capability.
The *RST selection is OFF.
SENSe[1]:LOGGing:SQUelch <boolean>
This command controls the logging squelch command. When enabled, further output of
logged text is inhibited if triggered for ten consecutive seconds.
The *RST selection is OFF.
SENSe[1]:LOGGing:ALARms <boolean>
This command controls the output of alarm conditions.
The *RST selection is OFF.
logging