Specifications
5-39
System Command Reference Section
Error Detector Measurement Functions
FETCH[:SENSe[1]]:GATE:ELAPsed? <NR3>
This node returns information about the degree to which the gating period has progressed.
If GATE:MANNer TIME is selected, then this command returns the elapsed time into the
gating period in units of seconds. If GATE:MANNer ERRors is selected, then this
command returns the elapsed errors into the gating period. IF GATE:MANNer BITS is
selected, then this command returns the elapsed clock bits into the gating period.
FETCH[:SENSe[1]]:LTEXt?
This query command returns one line of log output. If a line of text is not currently
available, then the message No text currently available is returned.
Refer to Error Detector Result and Configuration Logging, page 5-45, for information
on the configuration of the logging functions using this query command while the
instrument is under remote control.
FETCH:SENSe2:FREQuency? <NR3>
This returns the current frequency of the signal on the clock input. This measurement is
independent of the gating period.
Superseded by SENSe2:FREQuency[:CW|:FIXEd]?
Retained for backwards compatibility with Agilent 71600B systems.
FETCH:SENSe2:BCOunt?
Returns the accumulated bit count since the start of gating.