Specifications
5-34
System Command Reference Section
Error Detector Error Location
Error Detector Error Location
The Error Location features of the error detector have three forms, all of which may be
used when the instrument has any RAM-based pattern configured, i.e. the functions are
not available when any one of the five pure PRBS patterns is selected. The three forms
are: Bit BER, Error Location capture and Block BER.
Whole Pattern BER, Bit BER and Block BER are mutually exclusive operating modes of
the error detector. When in Block BER mode the bit error and clock counting functions
underlying the query commands in Error Detector Measurement Functions on page 5-36
are altered in accordance with the block start and length parameters. Pattern
synchronization is also affected.
In general Block BER is not measured on every repetition of the pattern. The number of
times a pattern will be repeated when configuring the pattern generator and detector
depends on the pattern length and is defined by table 4-2 of the Operating Manual. Where
it is important to make Block BER measurements on every repetition of the pattern the
pattern length must be set to a multiple of 256 bits.
SENSe[1]:ELOCation ONCE
This initiates an error location capture measurement. This is an overlapped command.
SENSe[1]:ELOCation?
The boolean response to this query command is true during an error location capture, and
false at all other times.
SENSe[1]:ELOCation:BEADdress <numeric value>
This command configures the single bit BER “BIT: Error address”
The query form of this command will only return the last configured value of the bit error
address provided no bit error has occurred (while a RAM pattern was selected) since the
last error location capture operation was started, for example, using the
“SENSe1:ELOC ONCE” command. Where a bit error has been captured the query
command returns the bit error address of the errored bit. This address is automatically
configured as the bit error address for the single Bit BER measurements. Thus when the
gated single Bit BER measurements are of interest, one would expect to re-start gating
after any error location capture had fired.