Specifications
5-19
System Command Reference Section
Pattern Generator CLOCK OUT (inverted)
OUTPut2:TERMination <numeric value>
Enables the clock termination level to be selected as 0 Volts or −2 Volts.
See Handling Coupled Parameters,page2-11.
The *RST selection is 0 V.
OUTPut2:COUPling AC|DC
Permits the clock output coupling to be set to ac or dc.
The *RST selection is dc 0 V.
Pattern Generator CLOCK OUT (inverted)
In this section the SOURCE11:VOLT commands control the electrical levels at the
CLOCK OUT port. Refer to the previous section on page 5-18 for SOURCE11:VOLT and
OUTPUT11 commands which provide the controls for the CLOCK OUT port.
NOTE See Handling Coupled Parameters,page2-11.
SOURce11:VOLTage[:LEVel][:IMMediate][:AMPLitude] <numeric
value>
Sets the peak to peak value of the clock output, in units of Volts.
This command is allowed only when tracking is disabled.
The *RST selection is 500 mV.
SOURce11:VOLTage[:LEVel][:IMMediate]:HIGH <numeric value>
This is used to set the dc high output level, in units of Volts.
This command is allowed only when tracking is disabled.
The *RST selection is 0 V.
SOURce11:VOLTage:TRACK <boolean>
Enables the Clock, Clock Outputs to operate independently or track together.
The *RST selection is ON.