Specifications
3-8
Interrogating the Instrument Status
Standard Event Status Register Group
Standard Event Status Register Group
The Standard Event Status register group is a 16-bit register group which gives general
purpose information about the instrument. It is unique in that it is the only reporting
register group programmed using SCPI common commands while the other reporting
registers are programmed using the SCPI Status command set.
NOTE This register conforms, in part, to the generalized status register model.
It comprises of an Event and Enable register, but no Condition register
or Transition Filter. Therefore all positive (true) states occurring in this
register are latched.
Figure 3-4 Standard Event Status Register
The bits in the Standard Event Status register group are defined as follows:
Table3-2 StandardEventStatusRegister
Bit # Mnemonic Description Bit Value
0 OPC Operation Complete bit. 1
1- Notused
2 QYE Query Error bit. 4
3 DDE Device Dependent Error bit. 8
4 EXE Execution Error bit. 16
5 CME Command Error bit. 32
6 URQ Not used 64
7PON PowerOnbit. 128
8-15 - These bits are not used