Specifications

13-9
Troubleshooting
Clock Loss Troubleshooting
Clock Loss Troubleshooting
If the clock frequency shown on the display is incorrect or if the CLK LOSS indicator is lit
on either the Error Detector or Pattern Generator, suspect that one of the following is
faulty:
Clock Source module
Display.
NOTE The CLK LOSS indicator will be lit if the clock signal is typically
<-10 dBm.
Clock Source Output
Access the Clock Source setup on the Display, check that the Clock Source Output is set
to ON. Check that the frequency of the Clock source has been set within the range of the
system, 1 GHz to 12.5 GHz for Agilent 71612 option UHx systems. If correct, use an
Oscilloscope or Power Meter to check the output level is >10 dBm. If good, the Clock
Source is good.
Data Loss Troubleshooting
The DATA LOSS indicator is lit on the Error Detector when no data transitions have been
detected over a 1 ms period. Normally, if there is a loss of input signal the SYNCH and
ERRORS indicators will be lit. A loss of clock signal may also cause the DATA LOSS
indicator to light, see CLK LOSS Troubleshooting.
To troubleshoot the systems first check the data being applied to the Error Detector and
use known good cables - if still faulty then suspect the Error Detector.
NOTE The Error Detector DATA IN port is very sensitive and will trigger on
background noise.