Specifications

12-9
Error Messages
Non-Permanent Errors
543 Failed to turn on logging Attempt was made to initiate logging failed. Contact
Agilent Technologies.
544 Failed to turn off logging Attempt was made to turn off logging failed. Contact
Agilent Technologies.
545 Log buffer overflow Logged data has overflowed the internal buffer
because the instrument has not output the data at a
sufficient rate to either the printer or external controller.
546 Too many alarm changes to log Alarm logging has caused an overflow of the internal
buffer. Stabilize the EdClkLs, DataLs and SyncLs flags.
550 Gating i/p asserted. Clk-data
align aborted
The GATING INPUT has been asserted causing
external disabling of bit error measurements which is
incompatible with a clock to data input delay eye-
centering operation.
551 Gating i/p asserted. 0/1 centering
aborted
The GATING INPUT has been asserted causing
external disabling of bit error measurements which is
incompatible with a data input 0/1 threshold eye-
centering operation.
900…999 All errors in this range are a consequence of firmware
errors or compatibility. Contact Agilent Technologies.
Non-Permanent Errors, continued
Error No. Displayed Message Description