Specifications
11-29
Performance Tests
Error Detector Performance Tests
Clock/Data Align
1. Press the hardkey followed by the key.
2. Adjust the data delay slowly via the RPG control until sync loss occurs. This will be
highlighted on the HP 70004A display.
3. Press followed by the softkey.
4. Verify that the Delay value shown on the lower left of the display changes and sync is
achieved after 5 seconds.
Data Input Sensitivity
1. Connect the equipment as shown in Figure 11-16.
Figure 11-16
data output
DATA DELAY
input & eye
CLK-DAT ALIGN