Specifications
11-19
Performance Tests
Pattern Generator Tests
4. Press , use the RPG control to set the clock amplitude to its
maximum level (1 V).
5. Press Autoscale on the scope and select the DELTA markers ON.
6. Set Marker 2 to the trough of the signal and Marker 1 to the peak of the signal. Read
Marker 1 as the “High Output level” and 0V as the “Maximum amplitude”. Enter these
values in the Performance Test Record.
7. Press , use the RPG control to set the Hi-level to its minimum
(−1.5 V).
8. Press , use the RPG control to set the clock amplitude to its
minimum level (500 mV).
9. Set Marker 2 to the trough of the signal and Marker 1 to the peak of the signal. Read
Marker 1 as the “Low Output level” and 0V as the “Minimum amplitude”. Enter these
values in the Performance Test Record.
S/R CLK ampltd
S/R CLK Hi-level
S/R CLK ampltd