Specifications

11-18
Performance Tests
Pattern Generator Tests
Subrate Clock Out
1. Re-connect the equipment as shown in Figure 11-9 and set the clock output frequency
to 10 GHz.
Figure 11-9
2. Check that the output amplitude and offset can be adjusted within the specifications
given in Table 11-3.
3. Press , , use the RPG control to set the
Hi-level to its maximum (0V).
subrate outputs
S/R CLK Hi-level