Specifications

11-16
Performance Tests
Pattern Generator Tests
5. Activate the Autoscale function on the scope and set the DELTA T Markers to be ON.
6. Set the STOP Marker on the last rising clock pulse edge on the right hand side of the
scope display.
7. Using the RPG control, increase the delay to its maximum, observing the selected
rising edge move across the display.
8. Set the START Marker to the final position of the rising edge and record the DELTA T
value (clk/data delay) on the Performance Test Record.
9. Set the delay to 0 ps.
10.Set the STOP Marker on the first rising clock pulse edge on the left hand side of the
scope display.
11.Using the RPG control, decrease the delay to its minimum, observing the selected
rising edge move across the display.
12.Set the START Marker to the final position of the rising edge and record the DELTA T
value (clk/data delay) on the Performance Test Record.