Performance Tests
Pattern Generator Tests
Clock Out and Clock (inverted) Out Amplitude and Frequency
1. Connect the HP 70843 main CLOCK OUT ports to the oscilloscope as shown in
Figure 11-5. Press
Figure 11-5
2. Set the CLOCK and CLOCK outputs to track each other. Press then set
to ON.
3. Set the clock frequency to 10 GHz at 0 dbm.
4. Set the pattern generator Pattern Trigger to trigger on clock (select
and set to ).
INST PRESET
clock output
C/TRACK
trigger & setup
PG TRIG PAT CLK CLK