Specifications
10-8
Data Logging
NOTE Squelching status continues through an end of measurement period.
That is, if actively squelching when an end of measurement period occurs, then at the start
of the next measurement period the instrument will continue to squelch. Logging is
resumed after one trigger free second. A message similar to the following is logged at the
end of squelching.
1993-04-23 23:11:04 End of Squelching Printing
When squelching terminates the error count and error ratio are logged. Squelching has no
effect when the softkey is used.
Results Storage
An internal buffer is used to store results when the volume of results is greater than the
print speed. If additional results occur when the buffer is full, then new results are
discarded. The following message is logged each time the results buffer becomes full.
1993-04-23 16:12:44 Printing Results Buffer Overflow - Results Lost
Once the results buffer begins to empty, new results are then stored. No message is logged
when this occurs.
Results Storage when Logging is Switched On
When logging is switched ON the internal buffer used to store results is emptied.
Disabling logging inhibits further storage of results but does not erase previously stored
results.
LOG ON DEMAND