Specifications
10-6
Data Logging
End of Measurement Period Logging
The following sets of results can be logged at the end of the measurement period:
Selectable using the softkey.
• Logs Main Results plus Interval Results plus G.821 Analysis
• Logs the results currently part of the Users Page (except delta
error results).
End of Measurement Logging Trigger
At the end of a measurement period the triggers to initiate logging are:
•Always
• Error Count >0
• Error Ratio > threshold (threshold range 1.0 to 1.0E−20)
NOTE The three triggers are mutually exclusive and cannot be changed while
the softkey is set to
To Log Results at the End of the Measurement Period
The following procedure explains how to configure the Agilent 71612 error performance
analyzer to log results selected on the Users Page, at the end of the measurement period
when the error count >0. Results are logged to an GP-IB printer.
1. Configure the error performance analyzer to perform a measurement over a single or
manual gating period.
2. Select the hardkey.
3. Set the softkey to .
4. Switch ON the softkey (the key is ON when it is underlined).
5. Select , then .
6. Set the softkey to . This sets the error detector data logger
to log all the results selected on the Users Page.
7. Set the softkey to .
The error performance analyzer is now configured to log the Users Page results if, at the
end of the measurement period the error count is greater then zero.
LOG PRD FULLUSR
LOG PRD FULL
LOG PRD USR
LOGGING OFF ON ON
logging
LOGGING OFF ON OFF
LOG TO GPIB
trigger end prd LOG END ERRS>0 exit
LOG PRD FULLUSR USR
LOGGING OFF ON ON