Specifications

7-35
Softkey Menu Descriptions
Gating Menu
Gating Menu
Path
or ,
Description
The softkey gives access to a menu of softkeys which control the error detector
measurement (gating) period. The three gating modes (MANUAL, SINGLE and
REPEAT) are a one-of-three selection.
Gating after a Power Loss
After a power loss the error detector will attempt to regain sync for approximately
25 seconds.
If sync is regained within 25 seconds gating will restart immediately.
If after 25 seconds has elapsed sync has not been regained, gating is forced to start.
The Power Loss Seconds result displayed is the overall time that gating was lost, and thus
the time that the system was unable to make a measurement.
NOTE Neither the gating mode, nor the gating period may be changed while
gating is active.
The softkeys are as follows:
Starts the error detector gating, or stops the current gating action
and begins a new gating period.
Stops the current gating action.
The next three keys are mutually exclusive:
Configures the error detector to make measurements over a
gating period controlled by the user.
Configures the error detector to make measurements over one
gating period and then stop.
Configures the error detector to make repetitive measurements
whereby one gating period follows another immediately. There
is no deadtime between the end of one period and the start of the
next.
gating
MENU
gating
gating
gating
RUN GATING
STOP GATING
MANUAL
SINGLE
REPEAT