Specifications

6
Press the
V2
softkey, then position the V2 cursor to the baseline.
The
V value at the bottom of the display should be within the test limits of
table 4-4. If a result is not within the test limits, see "Troubleshooting the
Oscilloscope," on page 4–40.
7
Continue checking the voltage measurement accuracy with the
remaining lines in table 4-4.
Table 4-4
Voltage Measurement Accuracy
Volts/Div
setting
Power supply
setting
Test limits
HP 54600B, HP 54601B HP 54602 HP 54603B
5 V/Div 35 V 34.24 V to 35.76 V 34.24 V to 35.76 V 34.04 V to 35.96 V
2 V/Div 14 V 13.70 V to 14.30 V 13.70 V to 14.30 V 13.62 V to 14.38 V
1 V/Div 7 V 6.848 V to 7.152 V 6.848 V to 7.152 V 6.808 V to 7.192 V
0.5 V/Div 3.5 V 3.424 V to 3.576 V 3.424 V to 3.576 V 3.404 V to 3.596 V
0.2 V/Div 1.4 V 1.370 V to 1.430 V 1.370 V to 1.430 V 1.362 V to 1.438 V
0.1 V/Div 700 mV 684.8 mV to 715.2 mV 684.8 mV to 715.2 mV 680.8 mV to 719.2 mV
50 mV/Div 350 mV 342.4 mV to 357.6 mV 342.4 mV to 357.6 mV 340.4 mV to 359.6 mV
20 mV/Div 140 mV 137.0 mV to 143.0 mV 137.0 mV to 143.0 mV 136.2 mV to 143.8 mV
10 mV/Div 70 mV 68.48 mV to 71.52 mV 68.48 mV to 71.52 mV 68.08 mV to 71.92 mV
5 mV/Div* 35 mV 34.48 mV to 36.52 mV 34.24 mV to 35.76 mV 33.08 mV to 36.92 mV
2 mV/Div* 14 mV 12.48 mV to 15.52 mV 13.70 mV to 14.30 mV 12.08 mV to 15.92 mV
1 mV/Div** 7 mV 6.696 mV to 7.304 mV
* Full scale is defined as 80 mV for the 5 mV/div and 2 mV/div ranges on HP 54600B, HP 54601B, HP 54603B.
** 1 mV/div range only on HP 54602B. Full scale is defined as 16 mV .
8
Disconnect the power supply from the oscilloscope, then repeat
steps 1 to 7 for channel 2 (channels 2 to 4 on the HP 54601B and
HP 54602B). On the HP 54601B and HP 54602B, channels 3 and 4,
check the 0.5 V/div and 0.1 V/div range only.
Service
Verifying Oscilloscope Performance
4–9