Specifications
1
Set up the oscilloscope.
a
Press
Setup
, then press the
Default Setup
softkey.
b
Press
Voltage
, then press the
V avg
softkey.
c
Set the Volts/Div to the first line of table 4-4.
d
Adjust the channel 1 Position knob to place the baseline near
(but not at) the bottom of the display.
2
Press
Cursors
, then press the
V1
softkey.
3
Using the cursors knob, set the V1 cursor on the baseline.
If you are in an electrically noisy environment, it can help to place a shorting
cap on the input BNC connector when positioning V1.
4
Connect the power supply to the oscilloscope and to the multimeter,
using the BNC tee and cables.
5
Set the power supply output to the first line in table 4-4.
Service
Verifying Oscilloscope Performance
4–8