Specifications

9
Adjust the channel 1 high-frequency response for 1.5 minor division
of overshoot (6%).
10
Repeat steps 4 through 9 for channel 2 (channels 2 to 4 on the
HP 54601B and HP 54602B).
High-frequency pulse response adjustment locations
High-frequency pulse
response adjustments
Figure 4–4
Service
Adjusting the Oscilloscope
4–36