Specifications
If the
+
5.1 V measurement is out of tolerance, adjust the
+
5.1 V adjustment
on the power supply. The
±
15.75 V supplies are not adjustable and are
dependent upon the
+
5.1 V supply. If adjusting the power supply does not
bring all the voltages within tolerance, see "Troubleshooting the
Oscilloscope," on page 4–40.
Low Voltage Power Supply adjustment location
(top side of oscilloscope
+5.1 V adjust
Figure 4–2
Service
Adjusting the Oscilloscope
4–30