Specifications

HP 54603B
Performance Test Record
Date_______________________________________
Serial No. ______________________________________ Test by_____________________________________
Test Interval ____________________________________ Work Order No._______________________________
Recommended Next Testing _______________________ Temperature _________________________________
Output of dc calibrator Limits Result
4.990 V to 5.010 V _________
Voltage measurement accuracy
Range Reading Test Limits Channel 1 Channel 2
5 V/Div 35 V 34.04 V to 35.96 V _________ _________
2 V/Div 14 V 13.62 V to 14.38 V _________ _________
1 V/Div 7 V 6.808 V to 7.192 V _________ _________
500 mV/Div 3.5 V 3.404 V to 3.596 V _________ _________
200 mV/Div 1.4 V 1.362 V to 1.438 V _________ _________
100 mV/Div 700 mV 680.8 mV to 719.2 mV _________ _________
50 mV/Div 350 mV 340.4 mV to 359.6 mV _________ _________
20 mV/Div 140 mV 136.20 mV to 143.8 mV _________ _________
10 mV/Div 70 mV 68.08 mV to 71.92 mV _________ _________
5 mV/Div 35 mV 33.08 mV to 36.92 mV _________ _________
2 mV/Div 14 mV 12.08 mV to 15.92 mV _________ _________
Bandwidth Test Limits Channel 1 Channel 2
≤±
3 dB _________ _________
Horizontal
t and 1/
t accuracy
Reading Test Limits Results
Frequency 10 kHz 9.959 kHz to 10.04 kHz _________
Period 100
µ
s99.59
µ
s to 100.4
µ
s _________
Frequency 1 MHz 995.7 kHz to 1.004 MHz _________
Period 1
µ
s 995.7 ns to 1.004
µ
s _________
Frequency 50 MHz 49.25 MHz to 50.77 MHz _________
Period 20 ns 19.70 ns to 20.30 ns _________
Trigger sensitivity Test Limits Channel 1 Channel 2
Internal trigger 25 MHz at 0.35 divisions _________ _________
60 MHz at 1 division _________ _________
External
External trigger 60 MHz at 100 mV p-p _________
25 MHz at 50 mV p-p _________
H