Specifications
HP 54602B
Performance Test Record
Date_______________________________________
Serial No. ______________________________________ Test by_____________________________________
Test Interval ____________________________________ Work Order No._______________________________
Recommended Next Testing _______________________ Temperature _________________________________
Output of dc calibrator Limits Result
4.990 V to 5.010 V __________
Voltage measurement accuracy
Range Reading Test Limits Channel 1 Channel 2 Channel 3 Channel 4
5 V/Div 35 V 34.24 V to 35.76 V __________ __________ __________ __________
2 V/Div 14 V 13.70 V to 14.30 V __________ __________ __________ __________
1 V/Div 7 V 6.848 V to 7.152 V __________ __________ __________ __________
500 mV/Div 3.5 V 3.424 V to 3.576 V __________ __________ __________ __________
200 mV/Div 1.4 V 1.370 V to 1.430 V __________ __________ __________ __________
100 mV/Div 700 mV 684.8 mV to 715.2 mV __________ __________ __________ __________
50 mV/Div 350 mV 342.4 mV to 357.6 mV __________ __________ __________ __________
20 mV/Div 140 mV 137.0 mV to 143.0 mV __________ __________ __________ __________
10 mV/Div 70 mV 68.48 mV to 71.52 mV __________ __________ __________ __________
5 mV/Div 35 mV 34.24 mV to 35.76 mV __________ __________ __________ __________
2 mV/Div 14 mV 13.70 mV to 14.30 mV __________ __________ __________ __________
1 mV/Div 7 mV 6.696 mV to 7.304 mV __________ __________ __________ __________
Bandwidth Test Limits Channel 1 Channel 2 Channel 3 Channel 4
≤±
3 dB __________ __________ __________ __________
Horizontal
∆
t and 1/
∆
t accuracy
Reading Test Limits Results
Frequency 10 kHz 9.959 kHz to 10.04 kHz __________
Period 100
µ
s99.59
µ
s to 100.4
µ
s __________
Frequency 1 MHz 995.7 kHz to 1.004 MHz __________
Period 1
µ
s 995.7 ns to 1.004
µ
s __________
Frequency 100 MHz 97.08 MHz to 103.1 MHz __________
Period 10 ns 9.699 ns to 10.30 ns __________
Trigger sensitivity Test Limits Channel 1 Channel 2 Channel 3 Channel 4
Internal trigger 25 MHz at 0.35 divisions __________ __________ __________ __________
Channel 1 & 2 150 MHz at 1 division __________ __________
Channel 3 & 4 250 MHz at 1 division __________ __________
Option 005 voltage
measurement accuracy
Test Limits Channel 1 Channel 2
Amplitude peak-to-peak
≥
63.6 mVp-p __________ __________
H
4–25