Specifications
HP 54600B
Performance Test Record
Date_______________________________________
Serial No. ______________________________________ Test by_____________________________________
Test Interval ____________________________________ Work Order No._______________________________
Recommended Next Testing _______________________ Temperature _________________________________
Output of dc calibrator Limits Result
4.990 V to 5.010 V __________
Voltage measurement accuracy
Range Reading Test Limits Channel 1 Channel 2
5 V/Div 35 V 34.24 V to 35.76 V __________ __________
2 V/Div 14 V 13.70 V to 14.30 V __________ __________
1 V/Div 7 V 6.848 V to 7.152 V __________ __________
500 mV/Div 3.5 V 3.424 V to 3.576 V __________ __________
200 mV/Div 1.4 V 1.370 V to 1.430 V __________ __________
100 mV/Div 700 mV 684.8 mV to 715.2 mV __________ __________
50 mV/Div 350 mV 342.4 mV to 357.6 mV __________ __________
20 mV/Div 140 mV 137.0 mV to 143.0 mV __________ __________
10 mV/Div 70 mV 68.48 mV to 71.52 mV __________ __________
5 mV/Div 35 mV 34.48 mV to 35.52 mV __________ __________
2 mV/Div 14 mV 12.48 mV to 15.52 mV __________ __________
Bandwidth Test Limits Channel 1 Channel 2
≤±
3 dB __________ __________
Horizontal
∆
t and 1/
∆
t accuracy
Reading Test Limits Results
Frequency 10 kHz 9.959 kHz to 10.04 kHz __________
Period 100
µ
s99.59
µ
s to 100.4
µ
s __________
Frequency 1 MHz 995.7 kHz to 1.004 MHz __________
Period 1
µ
s 995.7 ns to 1.004
µ
s __________
Frequency 100 MHz 97.08 MHz to 103.1 MHz __________
Period 10 ns 9.699 ns to 10.30 ns __________
Trigger sensitivity Test Limits Channel 1 Channel 2
Internal trigger 25 MHz at 0.35 divisions __________ __________
100 MHz at 1 division __________ __________
External
External trigger 100 MHz at 100 mV p-p __________
25 MHz at 50 mV p-p __________
H
4–23