Specifications

Table 4-11 Signal Generator Frequency Setting
Selected
Channel
HP 54600B HP 54601B HP 54602B * HP 54600B
Channel 1 100 MHz 100 MHz 150 MHz 60 MHz
Channel 2 100 MHz 100 MHz 150 MHz 60 MHz
Channel 3 100 MHz 250 MHz
Channel 4 100 MHz 250 MHz
*1mv/div to 5 mv/div = 100 MHz.
b
Press
Autoscale
.
c
Decrease the output of the signal generator until there is 1 vertical
division of the signal displayed.
The trigger should be stable. If the triggering is not stable, try
adjusting the trigger level. If adjusting the trigger level makes the
triggering stable, the test still passes. If adjusting the trigger does not
help, see "Troubleshooting the Oscilloscope," on page 4–40.
5
Repeat steps 1 through 4 for channel 2 on the HP 54600B and
HP 54603B (channels 2 to 4 on the HP 54601B and HP 54602B).
Service
Verifying Oscilloscope Performance
4–20