Specifications

Set up the logic analyzer
1 Set up the Calibration menu.
a Select the Procedure field, then select Osc Out.
b Select the Signal field, then select Probe Comp.
2
Set up the Channel menu.
a Press the Chan key.
b
Select the Coupling field, then select 1M / DC.
c Move the cursor to the Probe field, then use the RPG knob to dial in 1:1.
Verify the AC CAL OUTPUT port
1 Using the BNC cable, connect channel 1 of the oscilloscope to the AC CAL OUTPUT
connector.
2 In the Calibration menu select Autoscale.
3 Press the Meas key. Verify that the waveform is approximately 0.8 Vp_p at
approximately 1.0000 KHz. Record the reading in the performance test record.
Testing Performance
To test the CAL OUTPUT ports (oscilloscope)
3–68