Specifications

Connect the logic analyzer
1 Using a 6-by-2 test connector, connect channel 0 of Pod 1 to the pulse generator
channel 1 output.
2 Using the SMA cable and the BNC adapter, connect the External Input of the pulse
generator to the Main Signal of the function generator.
Acquire the data
1 Enable the pulse generator channel 1 output (with the LED off).
2 Press the blue key, then press the Run key to select Run-Repetitive. Allow the logic
analyzer to acquire data for at least 100 valid runs as indicated in the pattern
statistics field. Observe the X to O time field and ensure the X marker to O marker
time is between 47.496 and 47.504 µs during the test.
3 When the logic analyzer has acquired more than 100 valid runs, press Stop. The Min
X-O field, Max X-O field, and the Avg X-O field in the logic analyzer Pattern Statistics
menu should read 47.50 µs. Record the results in the performance test record.
Testing Performance
To test the time interval accuracy (logic analyzer)
3–64