Specifications
2 Check the data pulse width. Using the oscilloscope, verify that the data pulse width
is 4.000 ns, +0 ps or −100 ps.
a In the oscilloscope Timebase menu, select Scale: 1.000 ns/div.
b In the oscilloscope Timebase menu, select Position. Using the oscilloscope knob,
position the data waveform so that the waveform is centered on the screen.
c On the oscilloscope, select [Shift] + width: channel 1, then select [Enter] to display the
data signal pulse width (+ width(1)).
d If the pulse width is outside the limits, adjust the pulse generator channel 1 width until
the pulse width is within limits.
Testing Performance
To test the single-clock, multiple-edge, state acquisition (logic analyzer)
3–55