Specifications

2 Disable the pulse generator channel 1 COMP (with the LED off).
3 Using the Delay mode of the pulse generator channel 1, position the pulses
according to the setup time of the setup/hold combination selected, +0.0 ps or 100
ps.
a
On the Oscilloscope, select [Define meas] Define Time - Stop edge: rising.
b In the oscilloscope timebase menu, select Position. Using the oscilloscope knob,
position the rising edge of the clock waveform so that it is centered on the display.
c
On the oscilloscope, select [Shift] Time, then select [Enter] to display the setup time
( Time(1)-(2)).
d Adjust the pulse generator channel 1 Delay until the pulses are aligned according the
the setup time of the setup/hold combination selected, +0.0 ps or -100 ps.
Testing Performance
To test the single-clock, single-edge, state acquisition (logic analyzer)
3–32