Specifications
2 Set up the Format menu.
a Press the Format key. Select State Acquisition Mode, then select Full Channel/4K
Memory/100MHz.
b Select the field to the right of each pod, then select ECL in the pop-up menu. Use the
knob to access pods not shown on the screen.
3
Set up the Trigger menu.
a Press the Trigger key. Select Modify Trigger, select Clear Trigger, then select All in
the pop-up menu.
b Select Count Off. Press Select again, then select Time in the pop-up menu. Select
Done to exit the menu.
c Select the field labeled 1 under the State Sequence Levels. Select the field labeled
"anystate," then select "no state." Select Done to exit the State Sequence Levels menu.
d Select the field next to "a," under the label Lab1. Type the following for your logic
analyzer, then press the Select key.
HP 1660C/CS – "00AA" HP 1662C/CS – "00AA"
HP 1661C/CS – "002A" HP 1663C/CS – "000A"
Testing Performance
To test the single-clock, single-edge, state acquisition (logic analyzer)
3–26