Specifications
Timebase
Range 1 ns/div to 5 s/div
Resolution 20 ps
Delay Pre-trigger Range
Time/div Setting Available Delay
1 µs to 5 s/div -8 x (s/div)
1 ns to 500 ns/div -4 µs
Delay Post-trigger Range
Time/div Setting Available Delay
100 ms to 5 s/div 2.5 ks
1 µs to 50 ms/div 33,500 x (s/div)
1ns to 500 ns/div 16.7 ms
Triggering
Trigger Level Range Within display window (full scale and offset)
Trigger Modes
Immediate Triggers immediately after arming
condition is met.
Edge Triggers on rising or falling edge of
any internal channel, count adjustable from
1 to 32,000.
Pattern Triggers on entering or exiting a
specified pattern of both internal
channels, count adjustable from 1 to 32,000.
Pattern Duration Triggers on pattern duration specified
across one or both channels. Modes are
"Present<", "Present>", and "Range."
Patterns longer than 20 ns but shorter
than 160 ms can be detected.
Pattern Occurrence Triggers on entering or exiting a
specified pattern across one or both channels.
Auto-Trigger If enabled, the module will self-trigger
if no trigger condition is found within
approximately 50 ms after arming.
Events Delay The trigger can be set to occur on the
nth edge or pattern, as specified by the
user. The number of events (n) can be
set from 1 to 32,000 events. Maximum
count frequency is 70 MHz.
Intermodule Arms another measurement module or
triggers the rear panel BNC.
External If enabled, the oscilloscope will trigger
on an ECL level signal (-1.3 V).
General Information
Supplemental Characteristics (oscilloscope)
1–11