Specifications

Product Regulations
Safety IEC 1010-1:1990+A1 / EN 61010-1:1993
UL3111
CSA-C22.2 No. 1010.1:1993
EMC This product meets the requirement of the European
Communities (EC) EMC Directive 89/336/EEC.
Emissions EN55011/CSIPR 11 (ISM, Group1,Class A equipment)
IEC 555-2 and IEC 555-3
Immunity EN50082-1 Code
1
Notes
2
IEC 801-2 (ESD)4kV CD, 8kV AD 1
IEC 801-3 (Rad.) 3V/m 1
IEC 801-4 (EFT) 1kV 1
1
Performance Codes:
1 PASS - Normal operations, no effect.
2 PASS - Temporary degradation, self recoverable.
3 PASS - Temporary degradation, operator intervention required.
4 FAIL - Not recoverable, component damage.
Supplemental Characteristics (oscilloscope)
Vertical (at BNC)
Vertical Sensitivity Range 4 mV/div to 10 V/div in 1-2-4 increments
(1:1 Probe)
DC Offset Range (1:1 Probe)
Vertical Sensitivity Available Offset
4 mV/div to 100 mV/div ±2 V
>100 mV/div to 500 mV/div ±10 V
>500 mV/div to 2.5 V/div ±50 V
>2.5 V/div to 10 V/div ±250 V
Probe Factors Any integer ratio from 1:1 to 1000:1
Maximum Safe Input Voltage 1 M: ±250 V [DC + peak ac (< 10 KHz)], CAT II
50 : ±5 VRMS
Channel-to-Channel Isolation DC to 50 MHz: 40 dB, 50 MHz to 250 MHz: 30 dB
General Information
Supplemental Characteristics (oscilloscope)
1–10