Specifications

To remove and replace the power supply
1 Using previous procedures, remove the following assemblies:
Handle
Rear Feet
Cover
Disk Drive Assembly
WARNING
Hazardous voltages exist on the power supply. To avoid electrical shock, disconnect the
power from the instrument before performing the following procedures. After disconnecting
the power, wait at least three minutes for the capacitors to discharge before continuing.
2
Lift the PCB locking pins out of the chassis.
3 Slide the power supply out far enough to reach the power supply cables, then
disconnect them from the power supply.
4 Slide the power supply the rest of the way out the side of the instrument.
5 Reverse this procedure to install the power supply.
Check that the following assemblies are properly installed before installing the power supply:
Monitor
Front Panel
Switch Actuator
CPU Board
To remove and replace the CPU board
1 Using previous procedures, remove the following assemblies:
Handle
Rear Feet
Cover
Disk Drive Assembly
Power Supply
2 Disconnect the CPU board interface cable from the acquisition board by pressing
down on the cable release tabs on the cable socket located on the acquisition board.
3 Disconnect the HP-IB and Input/Output cables from the CPU board.
4 Disconnect the front panel cable from the CPU board.
5 Disconnect the I\O board cable from the CPU board.
6 Slide the board out the side of the instrument.
7 Reverse this procedure to install the CPU board.
Check that the following assemblies are properly installed before installing the CPU board:
Monitor
Front Panel
Switch Actuator
Replacing Assemblies
To remove and replace the power supply
6–7