Specifications

17 In the Data Memory Test menu, select Run, then select Single. The test runs one
time, then the screen shows the results. When the test is finished, select Done.
To run a test continuously, select Repetitive. Select Stop to halt a Repetitive Run.
Go to step 19 to exit the test system.
18
For the HP 1660CP-series Logic Analyzers, Select Analy PV, then select Patt Gen in
the pop-up menu. In the Patt Gen menu select Clock Source Test.
You can run all tests at one time (except the Output Patterns routine) by selecting All Tests.
To see more details about each test, you can run each test individually. This example shows
how to run the Clock Source Test. The Vector Memory Test, Address Counter Test, and
Instruction Tests are run in a similar manner.
When the tests finish, the status for each test shows Passed or Failed. Select Done.
19
To exit the tests, press the System key. Select the field to the right of the Sys PV
field.
20 Select the Exit Test System.
If you are performing the self-tests as part of the troubleshooting flowchart, return to the
flowchart.
Troubleshooting
To run the self-tests
5–23