Specifications
Specifications (oscilloscope)
The HP 1660CS Logic Analyzers also include the following specifications:
Bandwidth
(*,1)
DC to 250 MHz (real time, dc-coupled)
Time Interval Measurement Accuracy
(*, 2)
±[(0.005% X ∆t)+
(2 x 10
−6
x delay setting)+150 ps]
DC Offset Accuracy
(*)
±(1.0% of channel offset + 2.0% of full scale)
Voltage Measurement Accuracy
(*, 3)
± [(1.5% of full scale + offset accuracy)
+ (0.008 x V/div)]
Trigger Sensitivity
(*)
DC to 50 MHz: 0.063 x full scale,
50 to 250 MHz: 0.125 x full scale
Input R (selectable)
(*)
1 MΩ: ±1%, 50 Ω: ±1%
* Specifications (valid within ± 10° C of auto-calibration temperature, excluding bandwidth−see note 1 for bandwidth specification.)
1. Upper bandwidth reduces by 2.5 MHz for every degree C above 35° C.
2. Specification applies to the maximum sampling rate. At lower rates, the specification is:± [(0.005% x ∆t)+(2 x 10
-6
x delay setting) +
(0.15 x sample interval)] for bandwidth limited signals (t
r
= 1.4 x sample interval). Sample interval is defined as
1
sample
rate
3. Digitizing level = (#vertical divisions)(
1
2
)(
1
LSB
), where LSB=2
# bits in ADC
Specifications (pattern generator)
There are no specifications for the pattern generator portion of the HP 1660CP-Series Logic
Analyzers.
General Information
Specifications (oscilloscope)
1–4