Set up the logic analyzer
11 Set up the Configuration menu.
a Press the Config key. At the pop up menu, select Scope Channel.
b Select the Input field, then select C1.
c Move the cursor to the Probe field, then use the RPG knob to dial in 1:1.
d Move the cursor to the V/Div field, then use the PRG knob to dial in 400 mV.
e Move the cursor to the Offset field. Set the offset to 0 by typing 0, then pressing the
Select key.
f
Move the cursor to the Coupling field, then press Select. Select 50Ω / DC.
g Move the cursor to the s/Div field, then use the RPG knob to dial in 5.00 ns.
2
Set up the Display menu.
a Press the Display key.
b Select the Mode/Arm field, then select Average.
c Move the cursor to the Average # field. Enter "8" in the front-panel keyboard, then
press Done.
d Select the Grid field and set it to On.
e In the Waveform selection menu, delete channel 2. If channel 1 is not inserted, insert
channel 1.
Testing Performance
To test the trigger sensitivity (oscilloscope)
3–92