Specifications

To test the trigger sensitivity (oscilloscope)
Testing the trigger sensitivity verifies the performance of the following specifications:
DC to 50 MHz: 0.063 x full scale (0.25 division)
50 MHz to 250 MHz: 0.125 x full scale (0.5 division)
Equipment Required
Equipment Critical Specifications Recommended
Model/Part
Signal Generator 50 MHz and 225 MHz, 30 - 80 mV RMS output HP 8656B Opt. 001
Cable BNC 48-inch HP 10503B
Adapter Type N (m) to BNC (f) HP 1250-0780
Set up the equipment
Turn on the equipment required and the logic analyzer. Let them warm up for 30
minutes if you have not already done so.