Specifications

Set up the logic analyzer
1 Set up the Configuration menu.
a Press the Config key. At the pop up menu, select Scope Channel.
b Select the Input field, then select C1.
c Move the cursor to the Probe field, then use the RPG knob to dial in 1:1.
d Move the cursor to the V/Div field, then use the PRG knob to dial in 4.00 V.
e Move the cursor to the Offset field. Set the offset to 0 by typing 0, then pressing the
Select key.
f
Select the Coupling field, then select 1M / DC.
CAUTION
Set the Channel Coupling field to 1M / DC or damage to the equipment will result.
g Move the cursor to the s/Div field, then use the RPG knob to dial in 500 ns.
2
Set up the Display menu.
a Press the Display key.
b Select the Mode field, then select Average.
c Move the cursor to the Average # field. Type 32 on the front-panel keyboard, then
press Done.
d Select the Grid field and set it to On.
e In the Waveform menu, delete channel 2. If channel 1 is not inserted, insert channel 1.
Testing Performance
To test the offset accuracy (oscilloscope)
3–78