Instruction manual
Chapter 1 Performance Tests
Agilent 53181A Performance Test Record (Page 3 of 4)
Assembly-Level Service Guide 1-73
1
Agilent 53181A Performance Test Record
(Page 3 of 4)
Complete Performance Tests
Test 4: Option 015/030/050 Channel 2 Frequency Sensitivity (Agilent 53181A Only)
Actual Reading Specification
100 MHz–1.5 GHz (Option 015 Only)
100 MHz–2.7 GHz (Option 030 Only)
2.8–3.0 GHz (Option 030 Only)
.2–5.0 GHz (Option 050 Only)
.2–12.4 GHz (Option 124 only)
1. __________
2. __________
3. __________
4. __________
5. __________
≤ −27 dBm
≤ −27 dBm
≤ −21 dBm
≤ −23 dBm
≤ −23 dBm
Test
Number Test Description Minimum
Actual Reading
Maximum
5 Option 015/030 Channel 2
Frequency Accuracy
(Agilent 53181A Only)
100 MHz to 1.5 GHz Test for
Option 015 (Agilent 53181A):
Channel 2
Channel 2
99.999,999,93 MHz
1, 499.999,999 MHz
1. __________
2. __________
100.000,000,07 MHz
1,500.000,001 MHz
5 100 MHz to 3.0 GHz Test for
Option 030 (Agilent 53181A):
Channel 2
Channel 2
99.999,999,93 MHz
2,999.999,998 MHz
3. __________
4. __________
100.000,000,07 MHz
3, 000.000,002 MHz
5 200 MHz to 5.0 GHz Test for
Option 050: 4,999.999,997 MHz 5. __________ 5,000.000,003 MHz
5 200 MHz to 12.4 GHz Test for
Option 124: 12,399.999,992 6. __________ 12.400.000,008 MHz