Instruction manual
Chapter 1 Performance Tests
Test 5: Option 015/030/050/124 Channel 2 Frequency Accuracy (Agilent
53181A Only)
Assembly-Level Service Guide 1-67
1
(NOTE: For Options 050/124)
8 Set the Agilent 8340B to output 5.0 GHz at −17 dBm.
Verify the Agilent 53181A Counter reads 5.0 GHz ±3 Hz.
9 Record the actual reading in the Performance Test Record (Test 5, Line 5).
(NOTE: For Option 124)
10 Set the Agilent 8340B to output 12.4 GHz at –17 dBm.
Verify the Agilent 53181A Counter reads 12.4 GHz ±8 Hz.
11 Record the actual reading in the Performance Test Record (Test 5, Line 6).
12 Disconnect the test setup.