Chapter 1 Performance Tests
Test 4: Option 015/030/050/124 Channel 2 Frequency Sensitivity
1-58 Assembly-Level Service Guide
1
Test 4: Option 015/030/050/124 Channel 2
Frequency Sensitivity
This test verifies the frequency range and sensitivity of the optional 1.5-GHz (Option
015), 3-GHz (Option 030), 5.0-GHz (Option 050), or 12.4-GHz (Option 124)
Channel 2 for the Agilent 53181A Frequency Counter.
Equipment
Agilent 8663A Synthesized Signal Generator (Not needed for Option 050/124)
Agilent 8340B Sweep Oscillator
Agilent 1250-0780 Type N-to-BNC Connector (Not needed for Option 050/124)
Agilent 10503A BNC Cable (2)
Agilent 10100C 50Ω Feedthrough
Counter Setup
1Cycle the POWER key to preset the Counter.
2 Set the Counter as follows:
Freq & Ratio, FREQUENCY 2