Instruction manual

Chapter 1 Performance Tests
Test 2: Channel 1 Frequency Sensitivity (Agilent 53181A Only)
Assembly-Level Service Guide 1-53
1
4 Sweep the frequency from 100 kHz to 100 MHz. For Option 060, sweep
frequencies from 100 kHz to 225 MHz.
The Counter should read frequencies from 100 kHz to 100 MHz at an input power
level of 21.0 dBm (20 mVrms). Option 060 Counters should read frequencies from
100 kHz to 225 MHz at 9.5 dBm (75 mVrms).
5 Record the actual reading in the Performance Test Record (Test 2, Line 1).
If you are testing a Counter with front-panel terminals (Standard), record the
reading on Line 1 of the Standard portion of the test record.
If you are testing a Counter with rear terminals (Option 060), record the reading on
Line 1 of the Option 060 portion of the test record.
NOTE DO NOT test the front terminals if rear terminals are installed. The front terminal
performance is not specified when the rear terminals are installed.
NOTE The remaining sensitivity procedures are NOT required for Counters that contain
Option 060 Rear Terminals; thus, this completes the sensitivity portion of the
Frequency Performance Test. Go to the Frequency Accuracy Test on page 1-55.