Chapter 1 Performance Tests
Test 2: Channel 1 Frequency Sensitivity (Agilent 53181A Only)
1-52 Assembly-Level Service Guide
1
100 kHz to 100 MHz Sensitivity for Agilent 53181A
1 Connect the equipment as shown in Figure 1-12 with the signal generator
connected to Channel 1 of the Counter.
Figure 1-12. Channels 1 Frequency Sensitivity Test Setup
2 Set the Agilent 8663A Signal Generator to output a 100 kHz, 35.0 dBm sine
wave signal (20.0 dBm for Option 060) to the Counter’s Channel 1 input.
3 In 0.1 dB steps, increase the power level until the Counter displays a stable
reading of 100 kHz.
The Counter should display 100 kHz when the signal generator level
is 21.0 dBm (20 mVrms). Counters with Option 060 Rear Terminals should display
100 kHz when the signal generator level is 9.5 dBm (75 mVrms).
HP 53181A
Counter
HP 8663A
Synthesized Signal Generator
N-to-BNC
Connecto
r
Output
Channel 1
50
Feedthrough
Ref. In
(rear panel)
Input
(rear panel)
BNC Tee