Chapter 1 Performance Tests
Test 1: Trigger Level (Agilent 53181A Only)
1-50 Assembly-Level Service Guide
1
7 In the Counter’s Channel 1 Trigger/Sensitivity menu, change the SLOPE to
NEG.
8 On the Counter, press Run key.
9 Set the Agilent 3325B DC OFFSET to +60 mV.
10 Now, decrement the offset on the Agilent 3325B by −1 mV until the Counter’s
Gate annunciator flashes and continue decrementing until the Counter displays
approximately 1 MHz.
Observe the offset value on the display of the Agilent 3325B.
Record the value__________ mV.
11 Add the lower peak voltage (−40mV) of the 80 mVp-p signal to the offset value in
step 10 (for example, 21 mV − 40 mV = −19 mV.)
This result is the lower hysteresis level.
Record the lower hysteresis level __________ mV.
12 Now, add the upper hysteresis value (recorded in step 6 )and lower hysteresis
value (recorded in step 11).
This is the trigger level; it should be 0.0 V ±15 mV.
13 Record the trigger level value in the Performance Test Record (Test 1, Line 1).