Instruction manual

Chapter 1 Performance Tests
Test 7: Peak Volts, Channels 1 and 2 (Agilent 53131A/132A Only)
Assembly-Level Service Guide 1-37
1
Test 7: Peak Volts, Channels 1 and 2
(Agilent 53131A/132A Only)
This test verifies the Peak Volts accuracy specification of the Agilent 53131A and
Agilent 53132A Universal Counters.
Equipment
Agilent 3325B Synthesizer
Agilent 10100C 50 Feedthrough
Agilent 10503A BNC Cable
Counter Setup
1Cycle the POWER key to preset the Counter.
2 Press Other Meas key until VOLT PEAKS 1 is displayed.
3 Set CHANNEL 1 input to:
50/1M, 50
X10 Attenuate, Off
100kHz Filter, Off
4 Set CHANNEL 2 input to:
50/1M, 50
X10 Attenuate, Off
100kHz Filter, Off