Chapter 1 Performance Tests
Test 5: Option 030/050/124 Channel 3 Frequency Sensitivity (Agilent
53131A/132A Only)
1-30 Assembly-Level Service Guide
1
Test 5: Option 030/050/124 Channel 3 Frequency
Sensitivity (Agilent 53131A/132A Only)
This test verifies the frequency range, and sensitivity of the optional 3-GHz and
5-GHz Channel 3 (Option 030/050/124) for the Agilent 53131A and the Agilent
53132A Universal Counters.
Equipment
Agilent 8663A Synthesized Signal Generator (not needed for Option 050)
Agilent 8340B Sweep Oscillator
Agilent 1250-0780 Type N-to-BNC Connector (not needed for Option 050)
Agilent 10503A BNC Cable(2)
Agilent 10100C 50Ω Feedthrough
Agilent 11500D N-N Cable (Option 050 only)
Agilent 1250-1250 N(M) to SMA(F)
Agilent 1250-0777 N(F) to N(F)
Counter Setup
1Cycle the POWER key to preset the Counter.
2 Set the Counter as follows:
Freq & Ratio, FREQUENCY 3