Instruction manual
Chapter 1 Performance Tests
Test 4: Channels 1 and 2 Frequency Accuracy (Agilent 53131A/132A
Only)
1-28 Assembly-Level Service Guide
1
Procedure
1 Connect the equipment as shown in Figure 1-8 with the signal generator
connected to Channel 1 of the Counter.
Figure 1-8. Channels 1 and 2 Frequency Accuracy Test Setup
2 Set the Signal Generator to output 1 MHz at −7 dBm.
3 On the Counter, press Run key.
Verify the Agilent 53131A Counter reads 1 MHz ±1 mHz.
Verify the Agilent 53132A Counter reads 1 MHz ±200 µHz.
4 Record the actual reading in the appropriate place in the Performance Test
Record (Test 4, Line 1 for Agilent 53131A or Test 4, Line 3 for Agilent 53132A.
H
P 53131A/132A
C
ounter
HP 8663A
Synthesized Signal Generator
N-to-BNC
Connect
or
Output
Channel 1
50Ω
Feedthrough
Ref. In
(rear panel)
Input
(rear panel)
Channel 2